HYDROLYSIS CHAMBER - EN 12447-2021
(Conforming to EN 12447:2021 – Hydrolysis Resistance Test for Geotextiles)
Product Description
The Hydrolysis Chamber is a precision-controlled high-temperature water immersion system designed to evaluate the hydrolytic resistance of geotextiles and related polymeric materials under accelerated ageing conditions, in accordance with EN 12447:2021. The system ensures uniform temperature distribution and controlled exposure conditions to simulate long-term degradation due to moisture and heat.
Applicable Standards
- EN 12447:2021 – Determination of resistance to hydrolysis in water
System Configuration
- Insulated stainless steel hydrolysis tank
- Digital temperature control system (PID based)
- Specimen holding racks (inert material)
- Circulation system for temperature uniformity
- Drain and overflow system
- Control panel with safety features
Technical Specifications
Temperature System
- Temperature Range: Ambient to 100 °C
- Operating Temperature: 80 ± 2 °C
- Temperature Accuracy: ±1 °C
- Temperature Uniformity: ±2 °C (as per EN requirement)
- Control Type: PID Digital Controller
- Display Resolution: 0.1 °C
Tank / Chamber
- Material of Construction: Stainless Steel-SS 304(SS 316 optional)
- Inner Tank Capacity: 50 Litres
- Insulation: High-density mineral wool
- Outer Body: Powder-coated mild steel / SS (Optional)
- Lid: Hinged insulated lid with handle
Water System
- Medium: Deionised water (as per EN ISO 3696)
- Water Circulation: Continuous circulation pump
- Water Level Control: Manual
- Drain Valve: Provided for easy discharge
- Volume Ratio Compliance:
- Water mass : specimen mass ≥ 30:1 (as per EN 12447)
Specimen Holding System
- Rack Material: Stainless Steel / PTFE coated
- Design: Multi-tier perforated racks
- Specimen Placement: Free circulation of water ensured
- Capacity: typical 20 specimens (Customizable)
Heating System
- Heater Type: Immersion heaters
- Heating Capacity: 3 kW
- Protection: Dry-run protection and over-temperature cut-off
Control Panel
- Digital PID temperature controller
- Timer: Programmable exposure duration (hours/days)
- Indicators: Power, heating, fault
Safety Controls:
- Over-temperature protection
- Low water level cut-off
- MCB / fuse protection
Optional Features
- Data logging system with USB/PC interface
- Automatic water filling system
- Calibration port for temperature validation
Test Conditions Capability
- Continuous operation at 80 °C for long durations (up to 112 days or more)
- Stable hydrolysis environment with minimal temperature fluctuation
- Uniform exposure for all specimens
Power Requirements
- Supply: 220–240 V, Single Phase, 50 Hz
- Connected Load: 3 kW
Accessories (Recommended)at extra cost
- Deionised water system
- Tensile testing machine (for post-ageing evaluation)
- Temperature calibration probe
- Specimen preparation tools