FLAME PHOTOMETER - 1382E-
Model Number : 1382
Microprocessor Flame Photometer (2F) with Sodium (Na) & Potassium (K) filters, up to 5-point calibration, direct readout in ppm & meq/l, auto filter selection, auto ignition, 20x4 line LCD display, Single Aspiration, RS232 interface PC software at extra cost, printer attachment facility supplied with compressor and other accessories for lab & medical use. Detail Specification as per following page. Note: Double distilled water and LPG gas cylinder is required at the time of demonstration of flame photometer.
TYPE
Introductions:
- Microprocessor Flame Photometer (2F)with Sodium (Na) & Potassium (K) filters, upto 5 point calibration, direct readout in ppm & meq/l, auto filter section, auto ignition, 20 x 4 line LCD display, Single Aspiration,RS232 interface PC software at other accessories for lab
Salient features:
- Microprocessor Based with Printer Interface
- Four Measurements in Single Aspiration
- Setup Storage Facility
- Sample Data Storage Facility
- Curve Calibration Programmability Using Maximum of 5 Standards
- Multiple Calibration Curve Saving Facility
- Data Processing Through Curve Fitting Techniques, Least Square, Quadratic
- Online Help Key
- Direct Results in ppm and meq
- Useful for Medical & Soil Testing
Pre-installation requireements :
- We request to you make necessary arrangement of prerequisites LPG gas cylinder with regulator, double distilled water, NACL LR grade, KCL LR grade, NA standard solution (Minimum 5 solution Viz. 2 ppm, 4 ppm, 6 ppm, 10 ppm or as per requirement) and K standard solution (Minimum One Viz. 10 ppm or as per requirement), power supply 230 volt, 50 Hz 25 Amp power plug to carry out the demonstration at your site
| General range | Na : 0-100 ppm, K : 0-100 ppm, Ca : 0-100 ppm, Li : 0-100 ppm |
|---|---|
| Sensitivity General | Na: 0.5 ppm, K : 0.5 ppm, Li: 0.5 ppm, Ca: 15 ppm |
| Filters | Na & K supplied (Ca & Li are optional at extra cost) |
| Resolution | 0.1 ppm |
| Reproducibility | < 1% coefficient of variation for 20 consecutive samples using 10 ppm Na set as maximum standard |
| Linearity | 2% error when 3 ppm Na/K and 5 ppm Li are set as maximum standard Curve Fit |
| Accuracy | ± 2% of FS |
| Display | 20 x 4 alphanumeric LCD with backlit |
| Average Time | In-built in software |
| Flame System | LPG and dry oil free air |
| Detector | Silicon photodiode |
| Calibration | Up to 5-Point calibration with curve fitting software |
| Gas Control | Adjustable with knobs |
| Ignition System | Auto ignition |
| automiser | Axial flow type |
| Printer | Provision for attachment of dot matrix printer or HP LaserJet / inkjet (non-windows) printers with centronics parallel interface |
| RS-232C | Available (PC software at extra cost) |
| Power | 230 V + 10% AC, 50 Hz |
| Dimensions | 365 x 245 x 220 mm (L x B x H) (Approx.) |
| Weight | Main Unit 9 Kg (Approximately) |
| Optional Accessories At Extra Cost | |
| Additional filter calcium (Ca), Lithium (Li) can be provided | Rs. 9000.00 Each. |
| Barium (Ba) Filter | Rs. 13800.00 Each. |
| RS 232 Professional PC Software for computer Interface | Rs. 21000.00 |