NABL Accredited Calibration Laboratory

Salient Features:

  1. Cleanliness : ISO Class 5 as per ISO 14644‑1
  2. Complete Cabinet in IS 316 grade SS construction with stain finish
  3. Polycarbonate front door
  4. Mini Pleat HEPA Filter‑ separator less (H‑14 rating)


Advantage:

EIE ‑ Vertical Laminar Airflow system allows operation in sterile and particle free conditions because of the continuous flushing of the working area by unidirectional, Vertical and ultra filtered airflow, which assures full product protection.


Material of construction:

The Complete Cabinet in IS 316 grade SS construction with stain finish


Blower Assembly:

DIDW type blower system, which consists of dynamically & statically balanced aluminum centrifugal impeller driven by an single phase, 1440‑RPM motor, enclosed in an PU coated GI casing suitably suspended in a pair springs & connected to the filter chamber through flexible canvas duct.


Documentation:

Working Manual & Warranty certificate will be provided

Test certificate for HEPA Filter will be provided


The following Tests will be carried‑out at our factory before dispatch with Calibration certificate:

  1. Laser Particle Count test to confirm the cleanliness level as per ISO 14644‑1
  2. DOP integrity test for HEPA Filter as per IES‑RP‑CC‑002
  3. Velocity profile for HEPA filter as per IES‑RP‑CC‑002
  4. (∆) Differential pressure measurement to confirm the HEPA filter condition



eieLAMINAR AIR FLOW-VERTICAL-S.S. 316 CONSTRUCTION-2 X 2 X 2 FT-MINI PLEAT HEPA FILTER-