Door Safety Features

Following door safety features are provided for maximum operator safety

1) Process-lock to prevent opening of the door during the process.

2) Process-lock to prevent opening of the door after successful completion of the test cycle.  


Insulation

The gap between the sterilization chamber and outer jacket will be heavily insulated with minimum of 50-mm thick R. B. Glass Wool material to avoid thermal leakage


Piping Package

All process piping in contact with the chamber will be fabricated from SS-304 grade material with argon welding.


Steam Accessories

The jacket and chamber are provided with an over pressure safety valve. The jacket is provided with a steam trap.


Surface Finish

The inside surface of the chamber and doors are finished to a surface finish better than 0.8 Ra value. All other non-contact parts are polished to a surface finish better that 180 grit.


Temperature and Pressure Gauge

  1. One TEMPERATURE SENSOR will be provided; - Fitted at drain outlet. Inputs/outputs of all sensors and electronic parts are monitored, processed and controlled by Microprocessor based PID Controller for accurate data processing.
  2. A pressure gauge is connected with the inner chamber to indicate the steam pressure.

Electrical Switchgear

The control cabinet to be provided with a main isolator for the three-phase supply and a separate switch for the control supply. The fascia of the control panel is provided with indication LED’s and switches for operation. Switchgear make : Schneider


Requirement at buyer end

Water supply provision for steam generator